LOGIQ V2/LOGIQ V1User Guide 1-9
Direction 5610736-100 Rev. 9
Before the system arrives (continued)
Figure 1-1. Example Plug and Outlet Configurations
1. 100-120 VAC, 10A
Plug and Outlet Configuration
2. 220-240 VAC, 10A
Plug and Outlet Configuration
NOTE: Country-specific power cords are currently available for
Argentina, Australia/New Zealand, China, Denmark, India/South
Africa, Switzerland, United Kingdom, Europe, the United States,
Israel, Brazil and Japan.
WARNING
To avoid risk of fire, the system power must be supplied from a
separate, properly rated outlet. See ‘Before the system arrives’
on page 1-8 for more information.
Under no circumstances should the AC power plug be altered,
changed, or adapted to a configuration rated less than
specified. Never use an extension cord or adapter plug.
To help assure grounding reliability, connect to a “hospital
grade” or “hospital only” grounded power outlet.