Probes and Biopsy
5-30 LOGIQ V2/LOGIQ V1 – User Guide
Direction 5610736-100
Rev. 9
Displaying the Guidezone (continued)
DANGER
Failure to match the guidezone displayed to the guide may
cause the needle to track a path outside the zone.
It is extremely important that when using the adjustable angle
biopsy guides, the angle displayed on the screen matches the
angle set on the guide, otherwise the needle will not follow the
displayed guidezone which could result in repeated biopsies or
patient injury.