Measurement and Analysis
Vivid S5/Vivid S6 User Manual 275
R2424458-100 Rev. 2
Acquisition
1. Create an exam, connect the ECG device and make sure
to obtain a stable ECG trace.
2. Acquire 2D grey scale cineloops of an Apical long axis
(APLAX) view, an Apical 4 chamber view and an Apical 2
chamber view.
3. Store the loop (must store).
Note: It is recommended to acquire all three apical views
sequentially in order to get similar heart rate in all views.
• The frame rate should be between 37 and 80 frames per
second. A high frame rate is recommended for high heart
rate.
• The scanner should be configured to store at least 100 ms
before and after each heart cycle,
- or -
• The scanner should be configured to store 3 loops or more.
• If the stored loop contains more than one heart cycle, the
analysis will be done on the second last heart cycle.
• The entire myocardium should be visible.
• A depth range that includes the entire left ventricle should
be used.
Starting AFI
1. Open an APLAX view and press MEASURE.
2. In the Measurement menu, select AFI. The View selection
menu is displayed (Figure 7-11).