Vivid S5/Vivid S6 User Manual 53
R2424458-100 Rev. 2
Preparing the unit for use
The Vivid S5/Vivid S6 ultrasound unit must operate within the
proper environment and in accordance with the requirements
described in this section. Before using the system, ensure that
the requirements are met.
Site requirements
Optimal operation of the unit can be obtained by implementing
the following requirements:
Power requirements
The Vivid S5/Vivid S6 ultrasound unit uses a separate power
outlet for 100 – 120 VAC or 240 VAC, 50–60 Hz.
Operating Environment
Ensure that there is sufficient air flow around the
Vivid S5/Vivid S6 ultrasound unit when installed or operated.
Environmental requirements
The Vivid S5/Vivid S6 ultrasound unit requires constant
maintenance of its operational environment. Different
temperature and humidity requirements are specified for
operation, storage and transportation.
WARNING
Operating the unit with the wrong voltage range causes
damages, voiding the factory warranty.
Table 2-1: Vivid S5/Vivid S6 temperature and humidity requirements
Requirement Temperature Humidity Air Pressure
Operational 10–40 °C 30–85% 700–1060 hPa
Storage -10–60 °C 30–70% 700–1060 hPa
Transport -10–60 °C 30–70% 700–1060 hPa