68 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Figure 2-9: Proper Gel bottle storage
Unit acclimation time
Following transport the unit may be very cold or hot. Allow the
unit to acclimate before being switched on. Acclimation will
take one hour for each 2.5
o
C increment when the unit’s
temperature is below 10
o
C or above 40
o
C.
Proper Gel bottle storage Improper Gel bottle storage
o
C
02.557.510354042.5
o
F
32 36.5 41 45.5 50 95 104 108.5
Hours 43210023
o
C
45 47.5 50 52.5 55 57.5 60
o
F
113 117.5 122 126.5 131 135.5 140
Hours 45678910