responding to radio frequency interference, in violation of the
European Union Medical Device Directive and FCC
regulations.
Do not use devices which intentionally transmit RF signals, for
example, cellular phones, transceivers, or radio controlled
products, in the vicinity of this equipment as it may cause
performance outside the published specifications. Keep the
power to these types of devices turned off when near this
equipment.
EMC performance
Portable and mobile radio communications equipment (e.g.
two-way radio, cellular/cordless telephones, wireless computer
networks), other than those supplied by GE, should be used no
closer to any part of this system, including cables, than
determined according to the following method:
Table 1-1: Portable and mobile radio communications equipment distance requirements
Frequency range: 150 kHz - 80 MHz 80 MHz - 800 MHz 800 MHz - 2.5 GHz
Calculation
Method:
d=[3.5/V
1
] square
root of P
d = [3.5/E
1
] square
root of P
d = [7/E
1
] square
root of P
Where: d= separation distance in meters, P = rated power of the transmitter,
V
1
=compliance value for conducted RF, E
1
= compliance value for radiated RF
If the maximum
transmitter power
in watts is rated The separation distance in meters should be:
5 2.6 2.6 5.2
20 5.2 5.2 10.5
100 12.0 12.0 24.0