Measurement and Analysis
Vivid S5/Vivid S6 User Manual 255
R2424458-100 Rev. 2
Cardiac measurements
2D Measurements
2D Length measurements
1. Generate the 2D image.
2. Press
FREEZE to stop the cineloop.
Alternative: Press
CALIPER on the con-
trol panel and press
CALIPER assignable.
3. Press MEASURE on the Control Panel.
4. Select Caliper in the Measurement Menu (see Figure 7-1).
5. Trackball the cursor to the start point of the measurement.
6. Press
SET to anchor the start point of the measurement.
See the Status bar
to get the next step
to perform.
7. Trackball the cursor to the measurement end point.
The current distance value is displayed in the
Measurement result table and is instantaneously updated
when moving the cursor.
The measurement
display color on the
2D image changes
from green to blue
after completion of
the measurement.
8. Press SET to anchor the end point of the measurement.
The measurement result is displayed in the Measurement
result table.
9. To assign a label to the measurement, see page 252.
10. Repeat steps 5 through 8 to make additional length
measurements.
The measurements
displayed on the 2D
image and the cor-
responding results
are numbered.
2D length measurement ratio
1. Generate the 2D image.
2. Press
FREEZE to stop the cineloop.
3. Press
MEASURE on the Control Panel.
Alternative: Press
CALIPER and DIST
RATIO assignable.
4. Select Dist. ratio in the Measurement Menu (see
Figure 7-1).
5. Perform two length measurements as described in steps 5
through 8 in the above section.
The measurement results including the ratio (%) of the two
measured lengths are displayed in the Measurement result
table.