Probes
Vivid S5/Vivid S6 User Manual 561
R2424458-100 Rev. 2
Coupling gels
In order to assure optimal transmission of energy between the
patient and probe, a conductive gel or couplant must be applied
liberally to the patient where scanning will be performed.
Coupling gels should not contain the following ingredients as
they are known to cause probe damage:
Methanol, ethanol, isopropanol, or any other alcohol-based
product
• Mineral oil
• Iodine
• Lotions
• Lanolin
• Aloe Vera
• Olive Oil
Methyl or Ethyl Parabens (para hydroxybenzoic acid)
• Dimethylsilicone
Returning/shipping probes and repair parts
US Department of Transportation and GE Medical Systems
policy requires that equipment returned for service MUST be
clean and free of blood and other infectious substances.
CAUTION
Probes for neuro surgical intra-operative use must NOT be
sterilized with liquid chemical sterilants because of the
possibility of neuro toxic residues remaining on the probe.
Neurological procedures must be done with the use of legally
marketed, sterile, pyrogen free probe sheaths.
3. After removing from the germicide, rinse the probe
following the germicide manufacturer's rinsing instructions.
Flush all visible germicide residue from the probe and allow
to air dry.
CAUTION
Do not use gels (lubricants) that are not recommended. They
may damage the probe and void the warranty.
WARNING
Do not apply gel to the eyes. If there is gel contact to the eye,
flush eye thoroughly with water.