Scanning Modes
152 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Octave
Enables quick switching over between Octave imaging and
corresponding fundamental frequencies, without rotating the
Frequency selection rotary knob.
Focus Pos
Changes the location of the focal point(s). A triangular focus
marker indicates the depth of the focal point.
Up/Down
Flips the M-Mode display 180 degrees.
Color maps
Displays a color map menu to optimize the greyscale
presentation. The menu enables an option from a list of
non-linear gray-curves or different colorized curves to be
selected. Use the trackball to point to a color map and press
SET to activate the desired map.
AMM
Anatomical M-Mode (option) is only available for cardiac
applications. See further information in "Anatomical M-Mode"
on page 154.
Compress
Controls the amount of contrast in the image. An index number
is displayed in the status window to indicate the relative level of
compression.
Reject
Adjust reject level. When this control is increased, low-level
echoes are rejected and appear darker in the image. An index
number is displayed in the status window to indicate the
relative level of rejection.