Scanning Modes
192 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
The configuration alternatives are:
TSI start: 60 ms, 80, 100, 120, AVO or Manual control
TSI end: AVC - 200 ms, AVC - 150 ms, AVC - 100 ms,
AVC - 50 ms, AVC, AVC + 50 ms, AVC + 100 ms,
AVC + 150 ms, AVC + 200 ms, MVO, MVO + 100 ms,
MVO + 160 ms, MVO + 200 ms, MVO + 260 ms or Manual
control.
Note: Manual adjustment of TSI start and TSI end markers
is available. Press
MORE and adjust TSI START and TSI END.
To store the modified marker settings, press
IMG STORE
and choose the configuration setting Manual control to
avoid automatic adjustment of the markers.
To configure TSI markers:
1. Press
CONFIG and select the category Measure.
2. In the Measure category, select the sheet Advanced.
3. In the Application specific parameters section adjust TSI
Start and TSI End parameters by selecting a new value
from the combo menu displayed upon selection.
Optimizing TSI
Use apical view when imaging and ensure that both LV
walls and opposing walls are in the view.
Activate TSI from an optimized TVI display.
Low time to peak values may be masked out with a different
color using the TSI Cutoff control.
Optimize 2D gain for clean chamber
ATO can be used
Clean ECG trace
Acquire one or more complete heart cycles
Default TSI start and end times are suggested
When analyzing TEE images where systolic velocities are
negative, the detection mode may be changed to "Time to
peak negative velocity" using the Invert control.