Probes
Vivid S5/Vivid S6 User Manual 553
R2424458-100 Rev. 2
Inspecting the probe
After each use
1. Inspect the lens, the probe housing and the cable
(Figure 11-4).
2. Look for damage that might allow liquid into the probe.
Before each use
1. Inspect the lens, the probe housing and the cable
(Figure 11-4).
2. Look for damage that might allow liquid into the probe.
3. Test the functionality of the probe.
Figure 11-4: Probe parts
CAUTION
If any damage is found, DO NOT use the probe until it has been
inspected and released for further use by a GE service
representative.
1. Housing
2. Strain relief
3. Seal
4. Lens