Measurement and Analysis
Vivid S5/Vivid S6 User Manual 247
R2424458-100 Rev. 2
About Measurement results display
Be aware of the following:
Measurement results display
By default the system always displays absolute values for
parameters measured in Doppler. This means that values
from above and below baseline will all be displayed as
positive results.
For Cardiac this behavior cannot be changed. For
non-Cardiac the Absolute Value setting can be turned off in
Config -> Meas/Text -> Advanced, by setting the attribute
Absolute Value to Off.
• Calculated parameters
For calculated parameters the system uses signed values
in calculation formulas, and displays the absolute value of
the result.