Probes
Vivid S5/Vivid S6 User Manual 545
R2424458-100 Rev. 2
4. Auto-freeze capability is disabled.
5. Lens temperature is monitored until it reaches equilibrium
(temperature change rate of less than 0.06 °C per minute).
6. a: Measurement uncertainty for probes with temperature
sensor: 0.3 °C
b: Measurement uncertainty and probe variation for other
probes: 1.6 °C
Probe orientation
Some probes are provided with a green light (LED) orientation
marking near their head (see Figure 11-1). Probes which do not
have a LED have an indentation (notch) for orientation on the
probe housing. This LED, or notch, corresponds with the V
mark on the scanning screen. The V mark indicates the
orientation of the probe to the scan.
Figure 11-1: Orientation marking on probe and on screen
1. LED
2. Notch
3. V-mark on screen: indicates the orientation of
the probe to the scan.