Probes
544 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Maximum probe temperature
Notes: Lens temperature measured under following conditions
per IEC 60601-2-37 Amd. 2:
1. Thermocouple was placed at the geometric center of the
lens. Thermal phantom made with tissue-mimicking
material as referenced in IEC60601-2-37 Amd 2: Annex
BB.
2. a: Thermal phantom at 23 +/- 3 °C for all probes.
b: Temperature rise is measured and added to 33 for
External-Use probes or 37 °C for Non External-Use
probes.
3. Probe placed upright in contact with above thermal
phantom.
Probe Max Temp
3S -RS 40.5
M4S-RS 42.8
5S-RS 40.8
6S-RS 39.5
7S-RS 37.7
10S-RS 36.1
8L-RS 41.0
9L-RS 40.4
12L-RS 36.6
i12L-RS 40.5
4C-RS 37.8
8C-RS 37.3
E8C-RS 40.7
6T/6T-RS 40.8
6Tc-RS 42.3
9T/9T-RS 42.1
P2D 35.3
P6D 36.0