550 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Activating the probe
When a probe is connected to the unit it is automatically
detected.
To select a probe and an application
1. Press PROBE on the control panel. A list of the connected
probes will pop up.
2. Trackball to the desired probe.
An application menu for the desired probe is then listed.
3. Trackball to the desired application
4. Press
SET to launch the application.
CAUTION
Make sure that the probe and application names displayed on the
screen correspond to the actual probe and application selection.
Check that the correct TI category is displayed (see "Thermal
Index" on page 26). TIB must be displayed when a fetal
application is selected.