Probes
Vivid S5/Vivid S6 User Manual 549
R2424458-100 Rev. 2
To connect a probe (small connector - type RS)
1. Hold the probe connector vertically with the cable pointing
upwards.
2. Push the connector locking handle to the right-most
position.
3. Align the connector with the probe port and carefully push
into place.
4. Push the connector locking handle to the left-most position.
To connect a probe (large connector - type OR)
1. Hold the probe connector vertically with the cable pointing
upwards.
2. Turn the connector locking handle horizontically.
3. Align the connector with the probe port and carefully push
into place.
4. Rotate the locking handle to the full vertical position to lock
in place.
CAUTION
Fault conditions can result in electric shock hazard. Do not touch
the surface of probe connectors which are exposed when the
probe is removed.
WARNING
Do NOT touch the patient and any of the connectors on the
ultrasound unit simultaneously, including ultrasound probe
connectors.
CAUTION
Take the following precautions with the probe cables:
Keep free from the wheels.
Do not bend cable sharply.