Measurement and Analysis
Vivid S5/Vivid S6 User Manual 373
R2424458-100 Rev. 2
Follicle measurements length, width,
and height
You can make left and right ovary follicle volume
measurements from three distances.
1. Select the right or left side by clicking Rt or Lt.
2. Select Follicle.
An active caliper displays.
3. Measure the 3 dimensions length, width, height of the
follicle as follows:
4. Make the first distance measurement:
To position the active caliper at the start point, move the
Trackball.
To fix the start point, press
SET.
The system fixes the first caliper and displays a second
active caliper.
To position the second active caliper at the end point,
move the Trackball.
A dotted line connects the measurement points.
To complete the measurement, press
SET.
The system displays the distance value in the Results
Window. After the first and second measurement, the
system displays an active caliper.
5. To make the second and third distance measurements,
repeat step 4 above.
The three distances will display in the Results Window.
After the third measurement, the system displays the
follicle volume calculation including previous
measurements in the Results window.
Endometrium thickness (Endo)
To measure the endometrium thickness, make one distance
measurement.
1. Select Endo.
An active caliper displays.
2. To position the active caliper at the start point, move the
Trackball.
3. To fix the start point, press
SET.