Stress Echo
216 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Note: The number of segments (WMS segment model), the
range of scoring values (WMS scoring legend) and the initial
scoring value (WMS initial scoring) may be configured in
Config/Meas Text/Advanced under the Cardiac M&A category.
Scoring acquired loops
1. After image selection (see page 213), press ANALYZE.
The Stress Echo Analysis screen is displayed (see
Figure 5-9).
To edit a score, se-
lect it and choose a
new score.
2. Trackball to a segment in one of the scoring diagrams and
press
SET.
The Score pop-up list is displayed (see Figure 5-10).
3. Trackball to a score.
Alternative: Press
the arrow heads at
the bottom of the
scoring diagram
(see Figure 5-9)
4. Press SET.
The score is displayed in the relevant segment area in the
diagram (see Figure 5-10).
5. Repeat steps 1 through 3 to score relevant segments.
6. Rotate the assignable
CHANGE PAGE to display next group
of images.
7. Repeat steps 1 through 3 to score relevant segments on the
new loops.
CAUTION
The wall motion scoring result is assigned to the stress level of
the image, but will not be updated if the image is moved to
another stress level in the protocol at a later time. Images should
be correctly placed in the protocol when performing wall motion
scoring.