Measurement and Analysis
258 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
The measurements
displayed on the 2D
image and the cor-
responding results
are numbered.
11. To assign a label to the measurement, see page 252.
12. Repeat steps 5 through 10 to make additional volume
measurements.
2D Depth measurements
The measurements described in this section enable depth
measurement from the probe to a selected point.
Note: This measurement is disabled in the factory default
configuration. See "Measurement package configuration" on
page 338 for details about on how to enable it through
configuration menu.
To perform a depth measurement:
1. Generate the 2D image.
2. Press
FREEZE to stop the cineloop.
3. Press
MEASURE on the Control Panel.
See the Status bar
to get the next step
to perform.
4. Press the assignable POINT to select the depth
measurement function.
5. Trackball the cursor to the position to measure.
The current distance from the probe is displayed in the
Measurement result table and is instantaneously updated
when moving the cursor.
The measurements
displayed on the 2D
image and the cor-
responding results
are numbered.
6. Press SET to anchor the point.
The depth value (cm) is displayed in the Measurement
result table.