Safety
24 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Protective earth (ground) Internal
Earth (ground) Internal
Equipotentiality: indicates terminal to
be used for connecting equipotential
conductors when interconnecting
(grounding) with other equipment.
Connection of additional protective
earth conductors or potential
equalization conductors is not
necessary in most cases and is only
recommended for situations involving
multiple equipment in a high-risk
patient environment to provide
assurance that all equipment is at the
same potential and operates within
acceptable leakage current limits. An
example of a high-risk patient would
be a special procedure where the
patient has an accessible conductive
path to the heart such as exposed
cardiac pacing leads.
Bottom of unit
Attention - Consult accompanying
documents: alerts the user to refer to
the user documentation when
complete information cannot be
provided on the label.
This marking on the control panel is
especially intended to alert the user
to consult the user manual for use
BEFORE operation of the system.
Various
CAUTION - Dangerous voltage: used
to indicate electric shock hazards.
Various
Label Purpose Location