Archiving
438 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Scanners TCP/IP settings
To be able to use the network functions when connected to a
hospital network, the scanner must have a proper network
address. Typically source for this information in the network
administrator.
1. Press
CONFIG and log on as administrator.
2. Select the Connectivity category and TCP/IP subgroup.
The TCP-IP subgroup is displayed.
Figure 9-28: TCP/IP setting
3. In the IP settings area enter:
The IP address for the scanner
The subnet mask for the scanner
The IP address for the Default Gateway
1. Computer name: device’s name of type
VIVID7-00nnnn or ECHOPAC7-00nnnn, where
“nnnn” is the system’s serial number. Do not
change the computer name.
2. IP settings: system IP settings
3. Remote archive setup: remote archive IP
address and name (EchoPAC PC or Image
Vault)
4. Save TCP/IP settings. The changes will be
effective after the system is rebooted.