Archiving
444 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Scanner’s connectivity configuration
The scanners TCP/IP settings must be configured as
described in "Scanners TCP/IP settings" on page 438.
In addition, to work against the DICOM server the following
information has to be entered in the scanner:
The DICOM server IP address
The DICOM server port number
The DICOM server AE title (the server’s name)
Typically source for this information in the network
administrator.
Setup of the DICOM server in the scanner’s configuration
management package
DICOM dataflow selection
1. Press
CONFIG and log on as administrator.
2. Select the Connectivity category and Dataflow subgroup.
The Dataflow sheet is displayed.
3. Select the DICOM dataflow to configure in the Name
pull-down menu (see Figure 9-29).
Figure 9-29: The Dataflow sheet