Report
Vivid S5/Vivid S6 User Manual 511
R2424458-100 Rev. 2
Inserting variable parameters in a statement
Variable parameters such as patient name, institution name,
measurement values etc. can be inserted in a statement as
tagged information.
To insert variable parameters in a statement:
1. Place the cursor at the required position in the Findings text
field (or Conclusion text field).
2. Press Insert parameter.
The Insert parameter window is displayed (see
Figure 10-11).
3. Browse and select the actual parameter to insert.
Note: for measurement values, select first the scanning
mode.
4. Press OK.
The selected parameter is inserted in the statement as a
tag (e.g. the {EF} tag refers to EF measurement)
Note: to display correctly in the report, the actual
parameter value must exist, e.g. if a measurement value is
included in a statement as a variable parameter, a
measurement value must exist for the current patient,
otherwise the parameter name is displayed.
Figure 10-11: Insert parameter window