Safety
Vivid S5/Vivid S6 User Manual 47
R2424458-100 Rev. 2
connect to Vivid S5/Vivid S6 and are used simultaneously,
must be less than or equal to the rated supply of the
VividS5/VividS6.
• There must be adequate heat dissipation and ventilation to
prevent overheating of the device.
• There must be adequate mechanical mounting of the
device and stability of the combination.
• Risk and leakage current of the combination must comply
with IEC/EN 60601-1.
• Electromagnetic emissions and immunity of the
combination must conform to IEC/EN 60601-1-2.
General precautions for installing an alternate off-board,
remote device or a network would include:
• The added device(s) must have appropriate safety
standard conformance and CE Marking.
• The added device(s) must be used for their intended
purpose having a compatible interface.
• Signal or mains isolation devices and additional protective
earth may be needed to assure compliance with IEC/EN
60601-1-1.
CAUTION
The connection of equipment or transmission networks other
than as specified in the user instructions can result in an electric
shock hazard or equipment malfunction. Substitute or alternate
equipment and connections requires verification of compatibility
and conformity to IEC/EN 60601-1-1 by the installer. Equipment
modifications and possible resulting malfunctions and
electromagnetic interference are the responsibility of the owner.