Figure 2-6: Peripherals/Accessory Connector Panel
1. Two isolated AC sockets for peripherals (Max. 250VA total)
2. LAN 10/100 Base-TX Ethernet network connector
3. Two interchangeable USB ports (digital printers or other peripherals)
4. Audio output
5. DVI video output connection to external display
WARNING
Accessory equipment connected to the analogue and digital
interfaces must be certified according to the respective IEC
standards (e.g. IEC 60950 for data processing equipment and
IEC 60601-1 (1988) for medical equipment). Any person
connecting additional equipment to the signal input part or
output part is configuring the medical system, and is therefore
responsible that the system complies with the requirements of
the system standard IEC 60601-1-1 (2000). If in doubt, consult the
technical service department or your local representative.
Do not touch the conducting parts of the USB or Ethernet cables
when connecting equipment to the unit.