EasyManua.ls Logo

Gendex GX CB-500 - Chapter 13 Product Information; Technical Specifications

Gendex GX CB-500
218 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13-1
G990710 30 June 2008
Chapter
13
Product Information
Technical Specifications
X-ray Source
Tube Voltage: 120 kVp (eff)
Tube Current: 3-7 mA
Voltage Wave Shape: Constant Potential
Focal Spot: 0.0197 inches (0.5 mm)
Duty Cycle: 15%
Source to Sensor distance: 28.1 inches (71.4 cm)
Source to Patient distance*:19.5 inches (49.53 cm) (center of rotation)
* The patient must be properly positioned in the Head Support
Positioner Mechanism for each patient for all applications in order
to have the focal spot to skin distance as large as possible.
Minimum Filtration (at 120 kVp(eff)) (mm of aluminum equivalent):
10 mm or greater
Maximum Rated Continuous Tube Operation:130 kVp @ 0.5 mA
Maximum Rated Pulsed Tube Operation: 130 kVp @ 1mA
NOTE: Leakage technique factors are measured at the maximum specified
energy.
Maximum Deviation: kV: + 5 kV
mA: +
10%
Timer: ± 5%
X-ray Beam Size: 13 cm height x 13 cm wide
PAN: 1 cm width x 13 cm height
(Automatically collimated not to exceed image detector readable area)
Image Detector: 12.8 cm width x 12.8 cm height)
Sensor Front Panel Attenuation Value: Less than 1mm of aluminum
equivalent

Table of Contents

Related product manuals