15-18
Gendex GXCB-500
TM
Service Manual
G990710 30 June 2008
An annual radiation survey may be typically required. This
survey is typically required to be submitted to the local
agency or government body.
A phantom or patients may be used for system training.
Employees of the facility may not be used for this training.
The system shall be registered with the local agency or
government body.
X-ray Tube Assembly
Imaging Sciences International utilizes the SXR 130-15-0.5 X-ray
Tube from Superior X-Ray Tube Company to manufacture our X-
ray head assemblies.
Nominal X-Ray Tube Voltage 120 kV
Max. Tube Current 7 mA
X-Ray Tube Nominal Anode Input Power 65 W
X-Ray Tube Maximum Anode Heat Content
(HU = Kvp x mA x Time in Seconds)
30,000 HU
X-Ray Tube Single Load Rating 120kV, 5mA
Max. X-Ray Tube Assembly Heat Content 120K HU
Max. Continuous Heat Dissipation of Tube Assy 65W
High Voltage Supply Requirements 120 VAC at 10 amps
Loading Factors Concerning Leakage Radiation:
CT Scan = 120KV 5mA 12 mS x 618 or 309
PAN Scan = 94KV 5mA
< = 25mR/hr @ 1m