8-20
PAN Phantom Test
PAN Phantom Test is used to validate the PAN scan data capture.
To perform a PAN Phantom Test:
1. Prepare the Bite Tip by inserting the narrow edges of the Bite
Tip down into the Bite Tip Holder uprights. Then turn the Bite
Tip a 1/4 turn to lock into place.
2. Insert the Phantom Platform and Bite Tip Holder into the
Positioning Block. The Bite Tip should rest on top of the
platform.
3. Place PAN Phantom on platform with balls facing up and top of
arch resting on Bite Tip.
4. Press the Alignment Light button on the Patient Alignment
Panel to display lasers. Use the Horizontal Laser to adjust the
height of the phantom as shown below. Use the Vertical Laser to
center the phantom on the platform.
Lasers are available while the Volume tab is selected (not PAN ).
5. Start test by selecting File > New Patient from the Main menu
or right click an existing image from database and select
Acquire New Scan.
6. From the Patient Information Panel, click Add.
PAN
PHANTOM
PHANTOM PLATFORM
BITE TIP HOLDER
BITE TIP
LASER
LINE