8-21
Enter PAN Test in the First Name field and QA in the Patient
ID field.
7. Click OK to close the Patient Information box.
8. Click the Pan tab in the Acquisition window and select,
Exposure: Large
9. Click the Capture button to start the test.
WARNINGWARNING
The X-ray device may be dangerous to the Patient and operator if
you do not observe and follow the operating instructions. Do not
operate this system unless you have received training to perform a
procedure.
10. The system moves approximately 1/4 rotation to the Home
Position and then displays the scan parameters.
11. Click OK to start the scan process.
12. Press Scan button on Control Box. An audible alarm is sounded
and the X-ray ON light is illuminated during radiation exposure.