7-3
Calibration
G990710 30 June 2008
Collimators Calibration
It is recommended to perform the Collimators Calibration annually
to ensure optimal image quality. This calibration is also necessary if
a mechancial adjustment is made to the Beam Limiter or if image
quality has degraded. The Panel Calibration must be performed prior
to this. The Collimator Calibration runs in both Portrait and
Landscape positions and takes less than 3 minutes to complete.
Perform Collimators Calibration:
7. After performing the Panel Calibration click the Calibrate
(middle) button in the Collimators field.
WARNINGWARNING
The X-ray device may be dangerous to the Patient and operator if
you do not observe and follow the operating instructions. Do not
operate this system unless you have received training to perform a
procedure.
A window is displayed “Starting Collimator Calibration
Remove all objects from the field of view and click OK to start
X-ray exposure.”
8. Click OK to start the X-ray exposure.
9. On the Control Box, press
Scan when prompted. An
audible alarm is sounded
and the X-ray ON light is
illuminated during
radiation exposure.
10. The scan starts in the
Landscape and continues
into the Portrait mode,
displaying three screen
shots in both modes.
11. Click OK when calibration
completes (less than 3 minutes)