6-8
Gendex GXCB-500
TM
Service Manual
G990710 30 June 2008
Laser Alignments
The scanner has two alignment lasers.
• Centerline Laser - located on the front of Overhead Gantry
• Crosshair Laser - housed inside the X-ray Source Assembly.
Do not stare into laser. Severe personal injury (blindness) may
result.
NOTE: Ensure that the Patient Chair Alignment has been completed
before beginning this procedure.
Adjust Centerline Laser
There are three laser adjustments;
• Right-to Left
• Angle
• Laser Line Sharpness.