6-14
Gendex GXCB-500
TM
Service Manual
G990710 30 June 2008
Head Holder Alignment
NOTE: Ensure that the Laser Alignments have been completed
before beginning this procedure.
1.
3
LASER
LINE
If head support is installed, loosen locking knob
and remove head support.
2. Slide head holder into place.
3. Place the Position Alignment tool between the
temple pads with the alignment mark facing front.
4. Press the ALIGNMENT LIGHT button on the
Operator Panel to turn on laser.
5.
5
5/32
Loosen screws underneath head holder with allen
wrench (5/32) and adjust the head holder so that the
laser aligns with the alignment mark.
6. When aligned, re-tighten screws.
7. Press the Push To Release lever to open arms and
remove the Position Alignment tool. Do not
manually force arms open.
8. Remove head holder and re-install head support.