8-11
Enter Water Calibration in the First Name field and Noise Test
Landscape in the Patient ID field.
8. Click OK to close the Patient Information box.
9. From the Volume tab select the following:
Size of Reconstruction Volume: Dia 8.5 - H 8.5 cm
Resolution: 0.4 voxel
The X-ray device may be dangerous to the Patient and operator if
you do not observe and follow the operating instructions. Do not
operate this system unless you have received training to perform a
procedure.
10. Click Capture to start the scan process.
11. Click OK at the prompt to continue.
12. Press the Scan button on the Control Box when prompted. An
audible alarm is sounded and the X-ray ON light is illuminated
during radiation exposure.
13. At the Contourline prompt, select No and Cancel.
14. Access the HU measurement image for evaluation by double
clicking Coronal View (bottom image, 2
nd
from the right).
15. In the top-right image of Coronal view, slide the Horizontal line
to the middle of the water height.