8.2.3.6 Selection calibration mode
Here you can set the calibration mode of a sensor.
GO Systemelektronik recommends using the default setting.
The values of the data points are the running average from the last 15 seconds
∗
.
A detailed description of the calibration can be found in:
Appendix B - Calibration of a pH sensor
Appendix C - Calibration of an oxygen sensor
The calibration of the two optional current outputs only happens with the AMS software.
∗
Strictly speaking, these are the last 15 measurement values, but since all connected sensors (in fact all but the ISA
spectrometer) are automatically set to a measurement interval of one second during calibration, it's the running av-
erage of the last 15 seconds.
<–
One point calibration y0=0
One point calibration y0=fix
Two point calibration
Sensor menu 8.2.3
Calibration mode one point calibration with with a calibration
curve through the origin.
The button is also a status indicator.
Calibration mode one point calibration with with a calibration
curve thats y0-value is set by the calibration factors of the
sensor.
The button is also a status indicator.
Calibration mode two point calibration
The button is also a status indicator.
Switches back to the sensor menu.
One point calibration y0=0
One point calibration y0=fix