10.6.2 Load Calibration Data
When you use a measuring volume for the first time, calibrate it first.
Import the corresponding calibration data into the software. If you pur‐
chased the measuring volume together with the complete ARAMIS sys‐
tem, the calibration data is already loaded.
Procedure:
1. Start the software.
2. Open the calibration dialog in workspace Set Up with Calibrate Sensor.
3.
Choose Calibration object ► Manage Calibration Objects ► Create
Cross (Coded).
4. Enter the data from the certificate of the calibration cross.
The calibration data is available in the software.
Info
If the data of the calibration object are already in the ARAMIS Professional
Software, select the corresponding calibration object in step 3.
NOTICE
If you enter wrong calibration object data into the software, the meas‐
uring system probably provides inaccurate measuring data.
▶
Make sure that you enter all calibration data correctly into the soft‐
ware!
Calibrate the Sensor
0000001503_002_EN_02-12-2019 Page 45 (56)