9 Sensor Warm-Up
Before you use the ARAMIS Adjustable system for a measurement or exe‐
cute a calibration, the sensor must warm up to its operating temperature.
After you switched on the sensor, a warm-up time of 20 minutes is
required.
Info
If you calibrate the sensor before the warm-up time is reached, the calibra‐
tion results are possibly not correct.
Info
If you use the sensor for measurements before the warm-up time is
reached, the measuring results are possibly not correct.