64 GMC-I Messtechnik GmbH
Sequence Parameter (as of firmware V1.3.0)
The default test sequences can be adapted to your applica-
tion or test standard via the sequence parameter. The entered
sequence parameter settings are only valid for the currently
selected switch position (A1 to A8 or AUTO) and are retained until
they are changed. Not all of the parameters are relevant, depend-
ing on the selected DUT classification (protection category etc.).
Sequence Parameter Meaning
Visual inspection (1) Visual inspection (standard):
on: activate
off: deactivate
Visual inspection 2
(IEC 60974-4)
Visual inspection, function test, welding units
on: activate
off: deactivate
Function test Function test:
on: activate
off: deactivate
Protective conductor resistance test
RPE Protective conductor resistance test:
on: activate
off: deactivate
RPE IP Protective conductor resistance at test socket:
Select test current IP: ±200 mA= / 200 mA~ /
Feature G01: 10 A~ / feature G02: 25 A~
RPE IP permanent
connection
Protective conductor resistance with permanent connec-
tion:
Select test current IP: ±200 mA= / 200 mA~ /
Feature G01: 10 A~ / feature G02: 25 A~
RPE as Protective conductor resistance test:
Execute as individual or multiple measurement.
Multiple measurement: Repeat testing of various conduc-
tive parts as often as desired, in the event that it’s not
clear as to whether or not all accessible, conductive parts
are connected to each other or to the protective conduc-
tor.
RPE measurement
duration
Protective conductor resistance test:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here.
Insulation resistance test
RINS PC I Insulation resistance tests for PCI:
on: activate
off: deactivate
RINS PC II Insulation resistance tests for PCII:
on: activate
off: deactivate
RINS PC I and II
(VDE 0701-0702)
(IEC 60974-4)
Insulation resistance tests for PCI and II:
on: activate
off: deactivate
RINS at AP Insulation resistance tests at application parts:
on: activate
off: deactivate
Measurement duration
RINS probe
Insulation resistance test via probe:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here. Default setting: 3 s
Measurement duration
RINS AP
Insulation resistance tests at application parts:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here. Default setting: 3 s
RINS pri./sec.
(VDE 0701-0702)
(IEC 60974-4)
Insulation resistance test between the primary and sec-
ondary sides of PCIII DUTs
on: activate
off: deactivate
RINS PC II as
(VDE 0701-0702)
(IEC 60974-4)
Insulation resistance test:
Execute as individual or multiple measurement.
Multiple measurement: Insulation resistance is measured
between short-circuited mains terminals (L-N) and acces-
sible, conductive parts which can be contacted with test
probe P1 and are not connected to the housing, repeat as
often as desired.
Measurement duration
RINS PC II
Insulation resistance test:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here. Default setting: 3 s
RINS sec./PE
(VDE 0701-0702)
(IEC 60974-4)
Insulation resistance test between the secondary side and
PE of PCIII DUTs:
on: activate
off: deactivate
Leakage current tests
Reverse polarity Leakage current tests:
On:
Measurements are conducted with both polarities.
Off: Measurement is only conducted with one/momentary polar-
ity.
IPE
(VDE 0701-0702)
(IEC 60974-4)
Protective conductor current:
on: activate
off: deactivate
IPE measurement type
(active)
(VDE 0701-0702)
Protective conductor current test (mains to test socket):
Measuring method: Direct or differential
IPE measurement duration
(VDE 0701-0702)
(IEC 60974-4)
Protective conductor current test:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here. Default setting: 3 s
IE
(IEC 62353)
Device leakage current test:
on: activate
off: deactivate
IE measurement type
(active)
(IEC 62353)
Device leakage current test
(mains to test socket):
Measuring method: Direct or differential
IE measurement duration
(IEC 62353)
Device leakage current test:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here. Default setting: 3 s
IT measurement type
(active)
(VDE 0701-0702)
Touch current test (mains to test socket):
Measuring method: Direct P1 or differential P1
The “Differential P1” method is only advisable in this case
if the device under test has ground connections which
cannot be disconnected for testing.
IT
(IEC 62353)
(IEC 60601)
Touch current test
on: activate
off: deactivate
IT as
(IEC 62353)
Touch current test:
Execute as individual or multiple measurement.
Multiple measurement: Various accessible, conductive
parts are contacted with test probe P1 in order to mea-
sure current flowing to the protective conductor via the
probe – repeat as often as desired.
IT measurement duration
(IEC 62353)
Touch current test:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here. Default setting: 3 s
IT welding circuit
(IEC 60974-4)
Touch current test at welding circuit:
on: activate
off: deactivate
IT PC II as
(IEC 60974-4)
Touch current test at welding circuit:
Execute as individual or multiple measurement.
IT PC II measurement
duration
(IEC 60974-4)
Touch current test at welding circuit:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here. Default setting: 3 s
IP AC
(IEC 60601)
Patient leakage current AC:
on: activate
off: deactivate
IP DC
(IEC 60601)
Patient leakage current DC:
on: activate
off: deactivate
IP measurement duration
(IEC 60601)
Patient leakage current:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here. Default setting: 3 s
Test conditions / fault conditions
IA
(IEC 62353)
Leakage current test at application part:
on: activate
off: deactivate
Measurement duration IA
AP
(IEC 62353)
Leakage current test at application part:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here. Default setting: 3 s
Connection and fuse tests
Short-circuit test L-N Short-circuit test between L and N
1
on: activate
off: deactivate
Short-circuit test LN-PE Short-circuit test between LN and PE1
1
on: activate
off: deactivate
Display test instructions
Test instructions which are not necessarily required for experi-
enced inspectors
on: activate
off: deactivate
Fuse test Testing the fuses:
Mains fuses, test probe fuse P1,
application part fuses
Sequence Parameter Meaning