Model 11-R
GRIMM AEROSOL TECHNIK GmbH & Co.KG Seite 52 von 83
G Output gravimetry factor. Can be changed in standby mode in the range from 0.1 to 2.0, incre-
ment 0.05.
H Output hours of operation
I Interval for the normal output and storage on the SD card. Can be changed in standby mode.
0 = 1 minute
1 = 5 minutes
2 = 10 minutes
3 = 15 minutes
4 = 30 minutes
5 = 60 minutes
6 = 6 seconds
J Output of the channel thresholds in µm
J Channel thresholds
Jc:
0.25
0.28
0.30
0.35
0.40
0.45
0.50
0.56
Jc;
0.65
0.70
0.80
1.0
1.3
1.6
2.0
2.5
Jc:
3.0
3.5
4.0
5.0
6.5
7.5
8.5
10.0
Jc;
12,5
15,0
17,5
20,0
25
30
32
L Output measuring location pre-selection (location number) and change in standby mode
^L Country setting (E or U) for date output on the dust monitor (standby mode only).
E = Europe
U = USA
M Output mean value and sample volume (count mode in the example that follows)
Comments: The display on the model 11-R has four lines (Mc: Mc; mc: mc;).
N Turn on mass display, turn off count mode