GRIMM AEROSOL TECHNIK GmbH & Co.KG Seite 53 von 83
R Run measurement. Start the measurement from standby mode.
S Stop. Switch to standby mode.
T Time output. The clock can be set in standby mode. If the minutes are changed, the seconds
are set to zero.
^T Timer operation for turning the dust monitor on and off automatically.
In this operating mode, turning the Mini-LAS 11-R on and off can be programmed with the fol-
lowing input:
- Month (Mon)
- Day
- Hour
- Minute (Min)
U Keyboard lock
U
U=0 Keyboard unlocked
U=1 Unable to go to standby mode
U=2 No keyboard operation possible
V Output version number of the device software
W Output filter weight and corresponding sample volume (weight)
^Y Power off: Turn off the dust monitor
Z Output and subsequent reset of the mean values and volume (zero)
@ Output of the serial number