Page 28 250 Technical Reference
SETUP DISPLAY PROMPTS
When accessing Full or Safe Setup the indicator will beep twice and then display the following:
FULL or SAFE (depending on setup access type)
SETUP
Software Version (i.e. V1.0).
Calibration Counter (i.e. C.00010)
and Setup Mode Counter (i.e. F.00002) if NTEP is enabled at OPTION:USE
If an access code has been configured, the “ENTER PASS” prompt will display and the setup
access code must be entered to gain access. Refer to Accessing the Setup Mode on page 27,
SAFE.PC (Safe Security Access code for Digital Setup) page 32 and FULL.PC (Full Security
Access code for Digital Setup) page 33 for more information.
The title of the first Group (i.e. BUILD) will then be displayed.
SETUP LOCK OUT
If an attempt is made to enter Full or Safe Setup using an incorrect access code, the indicator will
respond with the message “ENTRY DENIED” and then the user will be returned to the weigh
mode. An access code counter has been set so that only three failed attempts can be made to
access Full/Safe Setup. On the fourth attempt the user will be 'locked out' of Full/Safe setup.
Should this occur the “ENTER PASS” prompt will not display, but instead the “ENTRY DENIED”
message displays and returns the user to the normal operating mode.
To reset the access code counter, turn off the indicator and turn it back on. The access code
counter will be reset to zero (allowing the user to enter the correct access code).
E
E
X
X
I
I
T
T
T
T
H
H
E
E
S
S
E
E
T
T
U
U
P
P
M
M
O
O
D
D
E
E
To save settings, exit setup, and return to the normal weighing mode use one of the following
methods. The Model 250 and 250SS can use any of the three examples. The Model 250X only
allows Method 3.
• Method 1: Press and hold both the [POWER] and [f ] keys together for two
seconds.
• Method 2: Press and hold both the [POWER] and [ZERO] keys together for two
seconds.
• Method 3: Press the [ZERO] (GRP) key until End is displayed and then press
[TARE] (ITM).
The indicator will beep and then display the following:
SAVING
Software Version (i.e. V1.0).
Calibration Counter (i.e. C.00010)
Setup Mode Counter (i.e. F.00003) if NTEP is enabled at OPTION:USE The current weight will
then display.
Warning: If the power is interrupted while in setup (i.e. by disconnecting the power cable or
pressing the [POWER] key), unsaved settings will be lost.