Version 2.4 Page 35
BAUD (Serial Baud Rate)
The baud rate determines the serial data transmission speed.
Options are: 2400, 4800, 9600
Default: 9600
BITS (Serial Format Options)
The BITS parameter allows the data transmission bit pattern and interface to be changed. The
display will show the current setting in the form n81- where each character has a meaning as
shown below.
Options are:
N or O or E: Parity bit: (N) None, (O) Odd, (E) Even
8, 7: Number of data bits
1, 2: Number of stop bits
-, D: DTR handshake disabled or enabled
Default: n81-. (For most applications the default setting is applicable.)
ADDRES (Indicator Address)
Use this option to set the Model 250 address when operating with network communications.
Range 01 to 31
Default: 31
RST.CON (Reset Printed Consecutive Number)
Use this option to reset the printed consecutive number back to 1. The indicator will prompt with
Cont. N. Press <EDT> to change to Cont. Y and <OK> to continue. When Cont. Y has been
chosen the unit will display “DONE” to indicate that the operation has been completed.
SET.PTS OR CHECKW
Settings within this group configure the setpoint (250 & 250SS) or checkweighing (250X) system.
Refer to Setpoints page 49 for a detailed explanation of setpoints.
SRC (Weight Source)
The weight source determines the weight reading that is used for the setpoint comparisons.
Options are:
GROSS: Use gross weight readings only.
DISP: Use the displayed weight (gross or net).
OFF: Both setpoints are off.
Default: OFF
TARG.HI or Over (Target for Overweight)
This is the target for setpoint 1 (output 1). Weights above this target will cause the setpoint to
become active and output 1 to be driven. The indicator displays the over annunciator. Refer to
Status Annunciators page 22.
Range: -99999 to 999999
Default: 000000