TARG.LO or Under (Target for Underweight)
This is the target for setpoint 2 (output 2). Weights below this target will cause the setpoint to
become active and output 2 to be driven. The indicator displays the under annunciator. Refer to
Status Annunciators page 22.
Range: -99999 to 999999
Default: 000000
CLOCK
Items within this group set date and time related functions.
FORMAT (Date Format)
This sets the date format.
Options are:
dd.mm.yy
mm.dd.yy
Default: dd.mm.yy
YEAR (Set Year)
Range: 2000 to 2099
MONTH (Set Month)
Range: 01 to 12
DAY (Set Day)
Range: 01 to 31
HOUR (Set Hour)
Range: 00 to 23 (24-hour format)
MINUTE (Set Minute)
Range: 00 to 59
TEST
Items within this Group allow access to the testing routines for the Model 250. With these
routines the scale base output can be monitored and the inputs and outputs can be tested.
SCALE (Scale Base Test Display)
Verifies the correct load cell capacity and/or load cell wiring is used. It sets up the unit as a
simple test meter to measure the load cell signal output. The display reads in milliVolts-per-Volt,
factory calibrated to 0.1% worst case. When accessing this item, initially there should be no
weight on the scale. In OIML or NTEP modes, this display is only active for five seconds before
returning to the menu.