Page 30 250 Technical Reference
RES (Count-by Resolution)
Sets the resolution (Count-by) of the display. The resolution is the number by which the indicator
will count.
Options are: 1, 2, 5, 10, 20, 50 or 100
Default: 1
UNITS (Weighed Units)
Sets the units for display and printing.
Options are: (g) grams, (kg) kilograms, (lb) pounds, (t) tons, (o) ounces (250X only), ( ) none (i.e.
other units).
Default: lb
HI.RES (High Resolution x 10 mode)
Sets the indicator display weight at 10 times resolution. This is intended for test purposes but
may be used for non-trade weighing.
Options are: ON or OFF
Default: OFF
CABLE (4-Wire or 6-Wire)
Sets the load cell input to operate in 4-wire (auto sense) or 6-wire mode.
Options are: 4 or 6
Default: 6
OPTION
Items within this Group are used to configure the operating parameters of the scale.
USE (Scale Use)
This is where the basic use of the scale is set (legal for trade). This setting configures the
indicator for Industrial, OIML or NTEP operation. Refer to page 45 for more information.
Options are: INDUST (Industrial), OIML or NTEP
Default: INDUST
FILTER (Reading Average)
The Model 250 can average a number of consecutive readings when calculating the displayed
weight. This is used to reduce unwanted weight fluctuations caused by vibrations or dynamic
forces. High settings will stabilize the display at the expense of rapid response to sudden weight
changes.
Options are: NONE, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0 (time in seconds)
Default: 0.5 (seconds)
MOTION (Motion Detection)
Sets how much weight variation over a defined time period is allowed before the displayed weight
is considered to be unstable. This value is displayed as weight change (0.5 or 1.0 graduations)
per second. When set to OFF, the Motion Detection is ignored and ZERO, TARE and PRINT
actions are instantaneous.
Options: OFF, 0.5-1.0, 1.0-1.0 (graduations per second)