2. Panel settings
Set the APS-1102A as shown below. See “3.4 Main Operation Example”, for descriptors
of the settings methods.
Table2-1. Panel Settings for Operation Check
⇒ Check that the control panel MEASURE screen shows 100 Vrms as the AC output voltage.
Figure 2-3. Operation Check
⇒ If the normal values are not displayed, it may be due to an operation fault. Contact GW Instek or
our agent.
4. When you turn off the power, check the OUTPUT on/off LED is off. Then press the power
switch to the bottom () position. This shuts off the power supply and turns off the power
source unit.