Using Autoset
Background
The Autoset function automatically configures the
panel settings to the best viewing conditions, in the
following way.
Selecting the horizontal scale
Positioning the waveform horizontally
Selecting the vertical scale
Positioning the waveform vertically
Selecting the trigger source channel
Activating the channels
Procedure
1. Connect the input signal to
the oscilloscope and press
the Autoset key.
Autoset
2. The waveform appears in the center of the
display.
GDS-1000A Series User Manual
42
Undo option
Before Autoset After Autoset
Undoing the
Autoset
To undo the Autoset, press
Undo (available for a few
seconds).
Adjusting the
trigger level
If the waveform is still
unstable, try adjusting the
trigger level up or down by
using the Trigger Level knob.
LEVE L
Limitation
Autoset does not work in the following situation.
Input signal frequency less than 20Hz
Input signal amplitude less than 30mV
Running and stopping the trigger
Background
In the trigger Run mode, the oscilloscope
constantly searches for a trigger condition and
updates the signal into the display when the
condition is met.
In the trigger Stop mode, the oscilloscope stops
triggering and thus the last acquired waveforms
stay in the display. The trigger icon at the top of
the display changes into Stop mode.
Pressing the Trigger Run/Stop key
switches between the Run and Stop
mode.