APPENDIX 
121 
Probe Specifications 
GDS-1062A/1102A/1152A Probe 
 
Applicable model & 
probe 
 GDS-1062A 
GTP-060A-4*
Position x 10  Attenuation Ratio  10:1 
  Bandwidth  DC ~ 60MHz  DC ~ 100MHz 
 
Input Resistance  10MΩ when used with 1MΩ input 
 Input Capacitance 23pF approx.
  Maximum Input 
Voltage 
500V CAT I, 300V CAT II (DC+Peak AC) 
Derating with frequency 
Position x 1  Attenuation Ratio  1:1 
  Bandwidth  DC ~ 6MHz 
 
Input Resistance  1MΩ when used with 1MΩ input 
  Input Capacitance  128pF approx.  47pF approx. 
 
Maximum Input 
Voltage 
300V CAT I, 150V CAT II (DC+Peak AC) 
Derating with frequency 
Operatin
 Cond.  Temperature  –10°C ~ 55°C
 Relative Humidity ≤85% @35°C 
Safety Standard  EN 61010-031 CAT II 
 
Applicable model & 
probe 
 GDS-1152A GTP-150A-2* 
Position x 10  Attenuation Ratio  10:1 
 Bandwidth DC ~ 150MHz
  Input Resistance  10MΩ when used with 1MΩ input 
  Input Capacitance  17pF approx. 
 
Maximum Input 
Voltage 
500V CAT I, 300V CAT II (DC+Peak AC) 
Derating with frequency 
Position x 1  Attenuation Ratio  1:1 
  Bandwidth  DC ~ 6MHz 
 
Input Resistance  1MΩ when used with 1MΩ input 
 Input Capacitance 47pF approx. 
 
Maximum Input 
Voltage 
300V CAT I, 150V CAT II (DC+Peak AC) 
Derating with frequency 
Operating Cond.  Temperature  –10°C ~ 55°C 
 Relative Humidity ≤85% @35°C 
Safet
 Standard  EN 61010-031 CAT II 
* Note: GW Instek reserves the right to change the probe model type (GTP-060A-4, 
GTP-100A-4, GTP-150A-2) at anytime without notice for probe model types of similar 
specification. 
    GDS-1000A Series User Manual 
122 
EC Declaration of Conformity 
 
We 
GOOD WILL INSTRUMENT CO., LTD. 
No.7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan 
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69, Lushan Road, Suzhou New District Jiangsu, China 
declares that the below mentioned product 
GDS-1062A, GDS-1102A, GDS-1152A 
Are herewith confirmed to comply with the requirements set out in the 
Council Directive on the Approximation of the Law of Member States 
relating to Electromagnetic Compatibility (2004/108/EC) and Low 
Voltage Equipment Directive (2006/95/EC). For the evaluation 
regarding the Electromagnetic Compatibility and Low Voltage 
Equipment Directive, the following standards were applied:
◎ EMC 
EN 61326-1 : 
EN 61326-2-1: 
Electrical equipment for measurement, control and 
laboratory use –– EMC requirements (2006) 
Conducted and Radiated Emissions 
CISPR11: 2003+A1: 2004+A2: 2006 
Electrostatic Discharge 
IEC 61000-4-2: 2001 
Current Harmonic 
EN 61000-3-2: 2006 
Radiated Immunity 
IEC 61000-4-3: 2006+A1: 2007 
Voltage Fluctuation 
EN 61000-3-3: 1995+A1: 2001+A2 : 2005 
Electrical Fast Transients 
IEC 61000-4-4: 2004+Corr.1 : 2006+Corr.2 : 
2007 
------------------------- 
Surge Immunity 
IEC 61000-4-5: 2005 
------------------------- 
Conducted Susceptibility 
IEN 61000-4-6: 2003+A1: 2004+A2: 2006 
------------------------- 
Power Frequency Magnetic Field 
IEC 61000-4-8: 2001 
------------------------- 
Voltage Dips/ Interrupts 
IEC 61000-4-11: 2004 
◎ Safety 
Low Volta
ment Directive 2006/95/EC 
Safety Requirements 
IEC/EN 61010-1: 2001