CONFIGURATION
59
CONFIGURATION
The Configuration chapter describes how to
configure panel settings to make measurements
and observations suited to the application needs.
Acquisition
The acquisition process samples the analog input signals and
converts them into digital format for internal processing. You may
select the normal, average, or peak detect acquisition mode.
Selecting the acquisition mode
Procedure
1. Press the Acquire key.
Acquire
2. Select the acquisition mode
between Normal, Average
and Peak Detect.
Normal
Peak
Detect
Average
Range Normal
All of the acquired data is used to
draw the waveform.
GDS-1000A Series User Manual
60
Average
Multiple data is averaged to form a
waveform. This mode is useful for
drawing a noise-free waveform. To
select the number, press Average
repeatedly.
Average number: 2, 4, 8, 16, 32, 64, 128,
256
Peak detect
To activate the Peak detect mode,
press Peak-Detect. Only the minimum
and maximum value pairs for each
acquisition interval (bucket) are used.
This mode is useful for catching
abnormal glitches in a signal.
Peak detect effect
using the probe
comp. waveform
1. One of the probe
compensation waveforms
can demonstrate the peak
detection mode. Connect the
probe to the probe
compensation output.
2. Press the Utility key.
Utility
3. Press ProbeComp.
4. Press Wave Type and select
the
waveform.
5. Press the Autoset key. The
oscilloscope positions the
waveform in the center of
the display.
Autoset
6. Press the Acquire key.
Acquire