APPENDIX
117
APPENDIX
Fuse Replacement
Procedure
1. Remove the power cord and remove the fuse
socket using a minus driver.
2. Replace the fuse in the holder.
Ratings
T1A, 250V
GDS-1000A Series User Manual
118
GDS-1000A Series Specifications
The specifications apply when the oscilloscope is powered on for at
least 30 minutes under +20°C~+30°C.
Model-specific specifications
GDS-1062A Bandwidth (–3dB) DC coupling: DC ~ 60MHz
AC couplin
Bandwidth Limit 20MHz (−3dB)
Trigger Sensitivity 0.5div or 5mV (DC ~ 25MHz)
1.5div or 15mV (25MHz~60MHz)
External Trigger
Sensitivity
~ 50mV (DC~25MHz)
~ 100mV (25MHz~60MHz)
Rise Time < 5.8ns approx.
GDS-1102A Bandwidth (–3dB) DC coupling: DC ~ 100MHz
AC coupling: 10Hz ~ 100MHz
Bandwidth Limit 20MHz (−3dB)
Trigger Sensitivity 0.5div or 5mV (DC ~ 25MHz)
1.5div or 15mV (25MHz~100MHz)
External Trigger
Sensitivit
~ 50mV (DC~25MHz)
~ 100mV (25MHz~100MHz)
Rise Time < 3.5ns approx.
GDS-1152A Bandwidth (–3dB) DC coupling: DC ~ 150MHz
AC coupling: 10Hz ~ 150MHz
Bandwidth Limit 20MHz (−3dB)
Trigger Sensitivity 0.5div or 5mV (DC ~ 25MHz)
1.5div or 15mV (25MHz~150MHz)
External Trigger
Sensitivit
~ 50mV (DC~25MHz)
~ 100mV (25MHz~100MHz)
Rise Time < 2.3ns approx.