CONFIGURATION 
63 
Real time vs Equivalent time sampling mode 
 
Background 
The oscilloscope automatically switches between 
two sampling modes, Real-time and Equivalent-
time, according to the number of active channels 
and sampling rate. 
Real-time 
sampling 
Once sampled data is used to reconstruct a single 
waveform. Short-time events might get lost if the 
sampling rate gets too high. This mode is used 
when the sampling rate is relatively low (1GSa/s 
or lower). 
Equivalent-time 
sampling 
Multiple numbers of sampled data are 
accumulated to reconstruct a single waveform. 
ETS restores more waveform detail but takes 
longer to update the waveform. This mode is used 
when the sampling rate becomes higher than 
1GSa/s. The maximum equivalent-time sampling 
rate is 25GSa/s. 
    GDS-1000A Series User Manual 
64 
Display 
The Display section describes how to configure the display settings: 
drawing type, waveform accumulation, contrast adjustment, and 
grid settings. 
Selecting vector or dot drawing 
 
Procedure 
1.  Press the Display key. 
Display
 
Vectors 
The sampled dots are connected by 
lines. 
Accumulating the waveform 
 
Background 
Accumulation preserves the old waveform 
drawings and overwrites new waveforms on top 
of it. It is useful for observing waveform variation. 
Procedure 
3.  Press the Display key. 
Display