EasyManuals Logo

GW Instek GPP-3060 User Manual

GW Instek GPP-3060
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #140 background imageLoading...
Page #140 background image
GPP-3060/6030 User Manual
140
Example
ISET1?
Returns the CH1 output current setting.
Command
:SOURce[1|2]:CURRent[:LIMit]:STATe?
Function
Queries the current limit state.
Return
0: The current limit has not been reached.
1:The current limit has been reached.
Example
:SOURce2:CURRent:STATe?
Return the current limit state of CH2.
Note
At output off or Load Mode,The return value of is
0.
Command
:SOURce[1|2]:VOLTage < NRf >
Function
Sets the output voltage amplitude.
Description
<NRf>
0.000-MAX
Example
:SOURce2:VOLTage 5.321
Sets the output voltage to 5.321V for CH2.
Note
About CH3, If the settings are not 1.8V/ 2.5V/3.3V
or 5V, the existing output will not be affected and
errors message will occur, Please refer to
":SYSTem:ERRor?" .
Command
VSET<X>:<NR2>
Function
Sets the Voltage level.
Description
X 1= CH1, 2= CH2
<NR2>Decimal number, range 0-32.000V.
Example
VSET1:20.345
Sets the CH1 voltage to 20.345V.